Model |
BCU-01P220-AW |
Cooling method |
Thermoelectric cooling |
Cooling capacity(*1) |
220W |
Heating capacity(*1) |
500W |
Environmental standard |
Single phase 100VAC to 240VAC |
Max. consumption power |
600VA |
Liquid circulation method |
Closed circulating |
Circulation liquid(*4) |
Ethylene glycol(dilute concentration of 30% or less)(*2)
Purified water(Ion-exchange water)(*3) |
Pipework |
CPC coupling(Supplied with CPC coupling, Outer diameter of nipple 9.5mm) |
Tank volume |
500ml(MAX) |
Pump throughput(*5) |
Approx. 3 liters/min |
Pump head(*5) |
Approx. 7.5m |
Temperature control method |
PID control provided with auto-tuning function |
Control method |
Polarity reversion, linear voltage control |
Inner temperature sensor |
Resistence temperature detector JIS Pt100/JPt100 three-wire system |
Temperature setting range(*6) |
-10℃ to +40℃ |
Representational accuracy |
±0.1℃ |
Resolution |
0.1℃ |
Temperature control accuracy |
±0.03℃ |
Working temperature |
0 to 40℃ |
Working humidity |
85%RH or less, Free from condensation |
External input and output |
○Analog output(0-10V) 1(*7)
○Alarm contact output(AC100V 1.0A/DC24V 1.0A) 1
○Event contact output(AC100V 1.0A/DC24V 1.0A) 1
○Communication : RS-485 1
○Communication : USB2.0 1
|
Safeguard |
○Fuse<AC250V 10A>(Double pole on primary side)
○Alarm buzzer
○Self-diagnosis function
Upper temperature limit/Lower temperature limit/
Abnormal temp. at cooling jacket/Abnormal temp. at radiation fin/
Failure of power to thermoelectric module/
Decreased circulation liquid flow
|
Noise |
49dBA(MAX) |
Site of use |
Indoor only |
Environmental standard |
RoHS |
Outline dimensions (mm) |
W225×H374×D273 |
Weight(*8) |
12.5kg |